X-DRS Ceiling Standard

Ceiling-suspended X-ray system with detector.
Mechanical version 
With particularly smooth operation for effortless positioning
Tracking version
With electro-mechanical tracking of the X-ray tube when the table or wall stand bucky is moved
X-DRS CEILING STANDARD - Ceiling-suspended X-ray system with detector

Product images

Technical data

Ceiling Stand

Minimum focus-ceiling distance
830 mm
Vertical telescope travel range
1800 mm, motorized and manual movement
Longitudinal travel range
3500 mm (optional with ceiling rail extension: 5645 mm);
detent positions configurable during installation
Transverse travel range
2150 mm (optional with 4380 mm traverse track: 3500 mm);
detent positions configurable during installation
Length of transverse track 
3030 mm (optional 4380 mm)
Tube rotation range, horizontal axis
–120°/ +120°
detent positions at –90°; 0°; +90°
Tube rotation range, vertical axis
–154°/ +182°
detent positions at –90°; 0°; +90°; +180°
Front touch user interface
Status display, stand control, generator control
Tracking
Vertical motorized tracking of a wall bucky table
Vertical motorized tracking of tabletop position

Tube

Nominal voltage
150 kV
Focal spot values
0.6 / 1.0
Nominal powers
52 kW / 103 kW
respectively thermal anode reference power = 0 W, at 150 Hz
Heat storage capacities
600 kHU (anode)
2430 kHU (tube assembly)

Elevating Table

Tabletop (W x L)
80 cm (31.5’’) x 240 cm (94.5’’)
Table height (motorized adjustment)
54 cm to 93 cm ± 1.5 cm (21.3’’ to 36.6’’ ± 0.6’’)
Tabletop travel
Longitudinal: ± 48 cm (± 18.9’’);
Transverse: ± 14 cm (± 5.5’’)
Max. patient weight
280 kg (617 Ibs) in normal position, 200 kg (441 lbs) otherwise

Wall Stand

Vertical movement
34 cm to 173 cm (13.4” to 68.1”) above floor (center to center)
Tilting Bucky
Tiltable from –20° to +90° with 0°/90° detent

Generator

Power rating
60 kV: 1 mA to 1000 mA
100 kV: 1 mA to 650 mA
125 kV: 1 mA to 524 mA
150 kV: 1 mA to 433 mA
Exposure voltage
From 40 kV to 150 kV
mAs range
From 0.5 to 800 mAs graduated either in 33 fixed values of one or in 65 values of 1/2 exposure point
Maximum mAS
60 kV: 800 mAs
100 kV: 560 mAs
125 kV: 444 mAs
150 kV: 373 mAs
Dimensions (L x W x H)
1020 mm x 570 mm x 542 mm

Do you need more information about our products and services?

The actual product configuration depends on the selected product variant and may differ from country to country. The products and product features shown in illustrations may differ in individual details from the current product range. Changes serving technical progress and errors excepted.

EXAMION GmbH

Erich-Herion-Str. 37
70736 Fellbach
Germany

+49 (0)711 120002-0
info[at]examion[dot]com