X-DR L WiFi KM (2nd)

Technical data
Detection method
Indirect conversion, a-Si TFT
Indirect conversion, a-Si TFT
Scintillator
CsI
CsI
Dimensions (W x H x D)
460.2 mm x 383.7 mm x 15.9 mm (18.1“ x 15.1“ x 0.6“), equivalent to standard 14‘‘ x 17‘‘ cassette
460.2 mm x 383.7 mm x 15.9 mm (18.1“ x 15.1“ x 0.6“), equivalent to standard 14‘‘ x 17‘‘ cassette
Weight
2.6 kg (5.7 lbs)
2.6 kg (5.7 lbs)
Pixel matrix
1,994 x 2,430 pixel
1,994 x 2,430 pixel
Pixel pitch
175 µm
175 µm
Active area
35 cm x 43 cm (14‘‘ x 17‘‘)
35 cm x 43 cm (14‘‘ x 17‘‘)
A/D conversion
16 bit
16 bit
Spatial resolution
3.4 lp/mm
3.4 lp/mm
Image transfer time
3 s (wireless)
2 s (wired)
3 s (wireless)
2 s (wired)
Trigger mode
Manual mode
Manual mode
Housing material
Carbon fieber plate
Carbon fieber plate
Weight limit
Local load: 150 kg at Ø 40 mm
Uniform load: 300 kg effective image area overall
Local load: 150 kg at Ø 40 mm
Uniform load: 300 kg effective image area overall
Enviromental conditions
Operating
Temperature: 10 °C to 30 °C (50 °F to 86 °F)
Humidity: 35 % to 80 % (non-condensing)
Atmospheric pressure: 70 kPa to 106 kPa
Not operating
Temperature: -10 °C to 40 °C (14 °F to 104 °F)
Humidity: 20 % to 90 % (non-condensing)
Atmospheric pressure: 70 kPa to 106 kPa
Transport / Storage
Temperature: -20 °C to 50 °C (-4 °F to 122 °F)
Humidity: 20 % to 90 % (non-condensing)
Atmospheric pressure: 70 kPa to 106 kPa
Operating
Temperature: 10 °C to 30 °C (50 °F to 86 °F)
Humidity: 35 % to 80 % (non-condensing)
Atmospheric pressure: 70 kPa to 106 kPa
Not operating
Temperature: -10 °C to 40 °C (14 °F to 104 °F)
Humidity: 20 % to 90 % (non-condensing)
Atmospheric pressure: 70 kPa to 106 kPa
Transport / Storage
Temperature: -20 °C to 50 °C (-4 °F to 122 °F)
Humidity: 20 % to 90 % (non-condensing)
Atmospheric pressure: 70 kPa to 106 kPa
Battery
Type: Lithium ion capacitor
Charging time empty to full: < 30 minutes
Duration in standby status: approx. 20 hours
Type: Lithium ion capacitor
Charging time empty to full: < 30 minutes
Duration in standby status: approx. 20 hours
Capture console, consisting of
X-HW More Acquisition Station
X-HW More Monitor 22“
X-AQS Software
Software controls detector, image acquisition and optimization, connects dose area product measurement device.
Options:
1. X-AQS PACS,
2. DICOM Modality (Send/Store, Worklist),
3. X-AQS Modality
X-HW More Acquisition Station
X-HW More Monitor 22“
X-AQS Software
Software controls detector, image acquisition and optimization, connects dose area product measurement device.
Options:
1. X-AQS PACS,
2. DICOM Modality (Send/Store, Worklist),
3. X-AQS Modality
Optional
Monitor: X-HW More Touchscreen 19“, X-HW More Monitor Eizo MS230W,
X-HW More Monitor Eizo EV2450
Monitor: X-HW More Touchscreen 19“, X-HW More Monitor Eizo MS230W,
X-HW More Monitor Eizo EV2450
Regulatory requirements and standards
CE 2460
CE 2460
Warranty
24-month warranty
24-month warranty
You can download the technical data sheet of the X-DR L WiFi KM as a PDF file here.
Do you need more information about our products and services?
The actual product configuration depends on the selected product variant and may differ from country to country. The products and product features shown in illustrations may differ in individual details from the current product range. Changes serving technical progress and errors excepted.