X-DR Mobile 320

Technical data
Electrical Specifications
Mains connection
115 Vac / 230 Vac ±10%, standard monophase with ground conductor.
Automatic setting of the equipment according to mains voltage 50/60 Hz ± 5 Hz
115 Vac / 230 Vac ±10%, standard monophase with ground conductor.
Automatic setting of the equipment according to mains voltage 50/60 Hz ± 5 Hz
Absorbed current
10 A
10 A
Line compensation
Automatic
Automatic
Line resistance
<1 Ω @ 115/230 Vac
<1 Ω @ 115/230 Vac
Standard socket
16 A @ 230 Vac
16 A @ 230 Vac
Power supply cable
8 m
8 m
Mechanical Specifications
W x L x H (in transport position)
618 mm x 1366 mm x 1444 mm
618 mm x 1366 mm x 1444 mm
Transport handle height
949 mm
949 mm
Focus-ground distance (Z-axis)
417 mm ÷ 2092 mm
417 mm ÷ 2092 mm
Max. height
2333 mm
2333 mm
Rotation of the monobloc
around the arm axis (α swivel): ± 180°
around its axis (γ swivel): 151° (102° ahead, 49° back)
around the arm axis (α swivel): ± 180°
around its axis (γ swivel): 151° (102° ahead, 49° back)
Movement
manual
manual
Weight
170 kg
170 kg
Environmental Conditions
Operating
Temperature: 10 °C to 40 °C (50 °F to 104 °F)
Humidity: 30 % to 75 % (non-condensing)
Atmospheric pressure: 700 hPa to 1060 hPa
Transport / Storage
Temperature: -25 °C to 70 °C (-4 °F to 122 °F)
Humidity: 10 % to 90 % (non-condensing)
Atmospheric pressure: 500 hPa to 1060 hPa
Operating
Temperature: 10 °C to 40 °C (50 °F to 104 °F)
Humidity: 30 % to 75 % (non-condensing)
Atmospheric pressure: 700 hPa to 1060 hPa
Transport / Storage
Temperature: -25 °C to 70 °C (-4 °F to 122 °F)
Humidity: 10 % to 90 % (non-condensing)
Atmospheric pressure: 500 hPa to 1060 hPa
X-ray Specifications
Nominal power (IEC 60601-1)
32 kW @100 kV, 320 mA, 100 ms
32 kW @100 kV, 320 mA, 100 ms
kV variation range
40 ÷ 125 kV at steps of 1 kV
40 ÷ 125 kV at steps of 1 kV
mA variation range
50 ÷ 400 mA
50 ÷ 400 mA
mAs variation range
0.1 ÷ 220 mAs
0.1 ÷ 220 mAs
Exposure times range
0.001 ÷ 2.2 s according to mAs
0.001 ÷ 2.2 s according to mAs
X-ray group
Monoblock
Max. power (100 kV - 320 mA): 32 kW
Max voltage at the tube: 125 kV
Max current at the tube: 450 mA
Min. inherent filtration @75kV: 1.1 mmAl
Thermal capacity available: 500 kJ
Continuous thermal dissipation: 55 W
Weight: 19.4 kg
Max. power (100 kV - 320 mA): 32 kW
Max voltage at the tube: 125 kV
Max current at the tube: 450 mA
Min. inherent filtration @75kV: 1.1 mmAl
Thermal capacity available: 500 kJ
Continuous thermal dissipation: 55 W
Weight: 19.4 kg
X-ray tube
Rotation speed: 3000 min-¹
Nominal High voltage: 130 kVp
Nominal focus dim. (IEC 60336): 0,8 mm small focus, 1,3 mm large focus
Nominal anodic power (IEC 60613): 16 kW small focus, 32 kW large focus
Thermal capacity of the anode: 80 kJ (107 kHU)
Max continuous anode dissipation: 300 W
Min. inherent filtration (IEC 522): 0,7 mmAl eq.
Rotation speed: 3000 min-¹
Nominal High voltage: 130 kVp
Nominal focus dim. (IEC 60336): 0,8 mm small focus, 1,3 mm large focus
Nominal anodic power (IEC 60613): 16 kW small focus, 32 kW large focus
Thermal capacity of the anode: 80 kJ (107 kHU)
Max continuous anode dissipation: 300 W
Min. inherent filtration (IEC 522): 0,7 mmAl eq.
Collimator
Collimation: Manual with internal light source, multilayer, squared field.
Coverage of the field at 100cm FFD (SID): min 0 x 0 cm, max 43 x 43 cm
Lighting source: Clusters of high-brightness LED power.
Focal distance measurement: Retractable tape measure (max extension 3 m)
Rotation: ± 120°
Weight: 5.5 kg
Collimation: Manual with internal light source, multilayer, squared field.
Coverage of the field at 100cm FFD (SID): min 0 x 0 cm, max 43 x 43 cm
Lighting source: Clusters of high-brightness LED power.
Focal distance measurement: Retractable tape measure (max extension 3 m)
Rotation: ± 120°
Weight: 5.5 kg
Do you need more information about our products and services?
The actual product configuration depends on the selected product variant and may differ from country to country. The products and product features shown in illustrations may differ in individual details from the current product range. Changes serving technical progress and errors excepted.