X-DRS Ceiling Standard
The X-DRS Ceiling Standard is an extremely robust X-ray system, suitable for the entire range of radiological examinations in hospitals. The various settings for X-ray doses and generator operation are fully configurable to the corresponding requirements of the operator or the examination, ensuring adequate power reserves in all cases.
In the standard version, the X-ray system is equipped with electromotive tracking functions that assist the operator to obtain correct positioning prior to X-ray exposure. The X-ray tube itself is motorised and is moved into position by the tracking system when the table or wall unit is moved. The stitching function version offers the highest degree of motorised movement, allowing fully automatic full-leg or full-spine imaging acquisition at the wall unit.
- Motorised tracking of the X-ray emitter when the table height changes, or when the bucky tray is moved on the wall stand
- Stitching option: Bucky movement automatically results in correct positioning for the required partial images
- An exclusively mechanical version is available for special applications
Technical Specifications
Minimum focus-ceiling distance | 830 mm |
Vertical telescope travel range | 1800 mm |
Longitudinal travel range | 3500 mm (optional: 5645 mm) Detent positions configurable during installation |
Transverse travel range | 2150 mm (optional: 3500 mm) Detent positions configurable during installation |
Length of transverse track | 3030 mm (optional 4380 mm) |
Tube rotation range, horizontal axis | –120° / +120° Detent positions at: -90°; 0°; +90° |
Tube rotation range, vertical axis | –154°/ +182° Detent positions at: -90°; 0°; +90°; +180° |
Front touch user interface | SID Horizontal tube rotation angle |
ACSS T | Vertical motorized tracking of a wall bucky table Electromagnetic brake for telescopic lift, transverse track 3 m long transverse track Vertical SID acquisition Integrated lock function, locks at any position along the ceiling track Telescope carriage with automatic stop at SID 115 cm |
Tabletop width | 80 cm (31.5’’) |
Tabletop length | 240 cm (94.5’’) |
Table height (motorized adjustment) | 54 cm to 93 cm ± 1.5 cm (21.3’’ to 36.6’’ ± 0.6’’) |
Tabletop travel | Longitudinal: ± 48 cm (± 18.9’’); Transverse: ± 14 cm (± 5.5’’) |
Vertical movement WS TL/TR | 34 cm to 173 cm (13.4” to 68.1”) above floor (center to center) |
Tilting Bucky | Tiltable from –20° to +90° with 0°/90° detent |
Power rating | 60 kV: 1 mA to 640 mA 100 kV: 1 mA to 550 mA 125 kV: 1 mA to 443 mA 150 kV: 1 mA to 366 mA |
Exposure voltage | From 40 kV to 150 kV |
Maximum mAs | From 0.5 to 800 mAs graduated either in 33 fixed values of one or in 65 values of 1/2 exposure point |
Max. mAS | 60 kV: 800 mAs 100 kV: 560 mAs 125 kV: 444 mAs 150 kV: 373 mAs |
Dimensions (L x W x H) | 1020 mm x 570 mm x 542 mm (40.2" x 22.4" x 21.3") |
Nominal powers | 52 kW/103 kW for focal spots 0.6/1.0, respectively thermal anode reference power = 0 W, at 150 HZ |
Heat storage capacities | 600 kHU (anode) 2430 kHU (tube assembly) anode cooling rate 162 kHu/min |
Voltage range | 150 kV, 150/180 Hz |
Questions and Answers
EXAMION will continue to stay a strong, reliable partner for the customers, even beyond the purchase process. We perform service, repair and maintenance of the X-ray system and carry out all required tests and measurements competently. Furthermore, we offer attractive financing models. Our sales representatives will be happy to advise you personally and according to your needs.
Yes, an image export of single images can be done in different ways by the users themselves. Available formats include jpg and DICOM. To export DICOM images, our X-AQS “Export Module” is required. This is usually part of the standard configuration of our X-ray software, so it should probably be installed on your system already as well. Depending on the configuration of the X-AQS software, the image data can also be stored on a specific path for further distribution or upload by cloud services. Please feel free to contact our representatives for more detailed information.
Yes, our experts are mostly able to accomplish this; however, we always have to check this individually. Therefore, you do not have to rely on the provider you want to switch from for the conversion.
The actual product configuration depends on the equipment selected and may differ from country to country. The products and product equipment shown in this illustration may differ in individual details from the current delivery program. Subject to changes in the interest of technical progress and errors excepted.